External dimensions (L x D x H) | 1433mm x 978mm x 1590-1673mm |
---|---|
Volume of air treated | 690m3/hr |
Sash opening type | Sliding Upward and Downward, Tempered Glass |
Voltage/Frequency | Single Phase, Connection to main power : 240 V – Internal circuit : 24V-DC/50-60 Hz |
Enclosure Material | Clear transparent and chemical resistant acrylic panel and electro-galvanised steel coated with anti-acid polymer |
Filtration Modules | Made of injected Polypropylene |
Compliance with applicable world safety standards | AFNOR NFX 15-211:2009: France – JG/T 385-2012 : China – SEFA 9 : USA – BS 7989 : England – DIN 12 927 : Germany – EN 1822 : 1998 (HEPA & ULPA) – CE Marking |
Face velocity meter | Built-in Electronic Hotwire anemometer continuously controlling and monitoring the air face velocity within 0.4 and 0.6 m/s. Compliance to DOSH, included LEV (LOCAL EXHAUST VENTILATION) Test Report |
Face velocity at sash opening | 0.4 to 0.6 m/s (80-120 fpm) according to international and local standard (USECHH Reg. 2000) |
Smart Software | eGuard Software : Monitor remotely the safety functions of the Erlab Ductless Fume Hood's performance, change of settings, and events history via computer software and/or Smartphone App. Able to connect up to 250 units of Green Fume Hood |
Enclosure containment | <0.05 ppm of SF6 detected at the maneque nose level measured according to ASHRAE 110 testing method |
Molecular filters efficiency | <1% of the PEL-TWA of the handled chemicals released downstream the filtration system |
Filter revolving technology | 2 identical molecular filters located below and above a fan/detection chamber ensuring no chemical escape when the lower layered filter reached its saturation. Upper layered filter will replace the saturated lower layered filter. Upper layer will be replaced with a new molecular filter. |
No. of Filter Column | 3 filter columns, with 6 Neutrodine filters |
Safety Alert System | Patented "Smart" technology consisting of safety alerts by LED pulsation and coded sounds to express the fan failure, insufficient face velocity or filter saturation/ breakthrough |
Tests Certificates and Reports | Intertek (NFX 15-211) |
Green Fume Hood GFH5300
Next General Laboratory Fume Hood
Engineered with a range of patented innovations, Green Fume Hood GFH5300 offers a safe, high performance, energy-efficient, fully flexible solution, ideal for virtually any environment form a cleanroom to a teaching laboratory. GFH5300 also in compliance to DOSH and International Standard, like NFX15-211, EN14175-3, ASHRAE 110-1995 and SEFA 9. Erlab Green Fume Hood offers a unique, modular filtration column and proprietary Neutrodine filtration technology, which can handle multidisciplinary tasks involving enerything from acids, solvents, liquids and powders while exhausting 100 times fewer contaminants than allowed by the official Threshold Limit Value (TLV)/ Permissible Exposure Limit (PEL).
Green Fume Hood can be left on at all times, sash up or down without consuming energy or polluting environment. At the same time, it maintains the DOSH’s standard air face velocity of 0.5 m/s or 100 fpm. Green Fume Hood also offers the first ever remote communication software, developed with microsoft technologies, to provide unprecedented management capabilities for a network of up to 250 fume hoods.
- Proprietary Neutrodine Filtratin Technology
- Exclusive twin-sensors saturation detection
- User identification and approval system (RFID card)
- Advance eGuard Communication Technology and Software
- Electronic filter identification system
- Revolving Filter System
- Adaptable Modular Filtration Column (AMFC)
- Visual and Audio Active Alert System