| CLEANLINESS IN WORK AREA | ISO CLASS 5, CLASS 100 (FED 209E) |
|---|---|
| HEPA FILTER | Filters provide more than 99.99% typical efficiency for particle size greater than 0.3 microns |
| AVERAGE AIR VELOCITY | 0.2~0.5m/s (adjustable) |
| SOUND EMISSION (dB(A)) | ≤62 |
| FLUORESCENT LIGHT INTENSITY (LUX) | ≤450 |
| RATED POWER (VA) | 450 |
| RMS | ≤5µM |
| EXTERNAL DIMENSIONS WITH BASE STAND (WxDxH) | 1500x760x1630mm |
| INTERNAL WORK AREA, DIMENSIONS (WxDxH) | 1380x650x510mm |
| NET WEIGHT/SHIPPING WEIGHT (KG) | 185/250 |


Heal Force AlphaClean 1300 Vertical Laminar Flow Cabinet
AlphaClean 1300 Vertical Laminar Flow Cabinet
Heal Force laminar flow clean benches set the standard in quality, design, and innovation that comes from a heritage of over 25 years of experience. We know how important it is to offer a high level of protection to operators, samples, and the environment with advanced technology. With an extensive trace record of safety, reliability and performance, Heal Force cabinets and benches make an idea investment for a wide range of applications.
Blower
Automatically compensates for normal power line variation, air diruptiona and filter loading. Motor consumes less energy, reduce heat output and operates more quietly.
Illumination lamp
Illumination provides sufficient brightness to theworking chamber.
UV lamp
Emission of 235.7 nanometers for most efficient decontamination. Automatic UV timer program saves your time during daily work
Satety grade glass
The front glass window provides protection from explosion and UV with more comfortable viewing
Work tops
Standard constructed of seamless, non-porous type 304 stainless steel
Prefilter
Chemically and thermally enhanced technology increase filtration capacity and efficiency,together with ergonomical filter change design for easy maintenence
Friendly communication
Centered and angled down color-display for better at-a-glance viewing of safety and performance data.
Long life HEPA filter
Long life HEPA filter provides high separability for particles. Self-compensation for the clogging of filters and life span indicator optimize working condition and minimizes services
Real-time airflow monitoring
Velocity of airflow and temperature are inspected and indicated on control panel.Alert the user if airflow is insufficient
Universal electrical outlets
Two standard electrical duplex receptacles, with ground fault interruption and splash covers.